Article Figures & Data
Figures
Figure 1. Aerosol delivery via high-flow nasal cannula with versus without exhaled humidity in adult manikin.
Tables
- Results of aerosol delivery via HFNC with different models and mouth status
Mouth status Flow setting(L/min) Inhaled dose with 2 model, % P NMT model Adult manikin Closed 15 24.7 ± 0.4 18.9 ± 1.4 <0.001 45 6.5 ± 0.9 5.2 ± 0.8 0.038 Open 15 21.1 ± 0.6 24.4 ± 1.3 0.001 45 7.7 ± 1.0 6.1 ± 0.5 0.013 P 15 0.0079 0.0079 45 0.087 0.0952