RT Journal Article SR Electronic T1 An Interrater Reliability Study of Pulmonary Function Assessment With a Portable Spirometer JF Respiratory Care FD American Association for Respiratory Care SP 665 OP 672 DO 10.4187/respcare.07116 VO 65 IS 5 A1 Hao Zhang A1 Li Li A1 Daozhen Jiao A1 Yanjie Yang A1 Chunhong Pan A1 Ling Ye A1 Sulan Wei A1 Meiling Jin YR 2020 UL http://rc.rcjournal.com/content/65/5/665.abstract AB BACKGROUND: In this study, we aimed to validate the agreement between pulmonary function measurements obtained with a portable spirometer and measurements obtained with conventional spirometry in Chinese pediatric and adult populations.METHODS: Pulmonary function testing was performed to evaluate subjects enrolled at Shanghai Zhongshan Hospital (n = 104) and Shanghai Children's Medical Center (n = 103). The portable spirometers and conventional devices were applied to each subject with a 20-min quiescent period between each measurement. Pulmonary function parameters of FVC, FEV1, peak expiratory flow, maximum expiratory flow at 25%, 50%, and 75% of FVC (MEF25, MEF50, and MEF75, respectively), and FEV1/FVC% were compared with intraclass correlation and Bland-Altman methods.RESULTS: A satisfactory concordance of pulmonary function was observed between spirometry measurements obtained with portable versus conventional spirometers. Intraclass correlation indicated excellent reliability (>0.75) for all pulmonary function indicators in pediatric and adult subjects. Significant positive correlations of all variables measured with different spirometers were observed (all P < .001). No significant bias was observed in either group, although limits of agreement varied. Funnel effects were observed for peak expiratory flow in pediatric subjects and for FVC, FEV1, MEF50, and MEF25 in adult subjects.CONCLUSIONS: The portable spirometer is an alternative to the conventional device for the measurement of pulmonary function. Compared with the conventional device, the portable spirometer is expected to provide convenient, operational, and financial advantages.