%0 Journal Article %A Rachel Culbreth %A Robert Brent Murray %A Kyle Brandenberger %A Douglas S Gardenhire %T Comparison of Portable Oxygen Concentrators and Inspired Oxygen Levels Using a Simulation of COPD %D 2021 %J Respiratory Care %P 3599013 %V 66 %N Suppl 10 %X Background: Portable oxygen concentrators (POCs) have shown efficacy in delivering adequate oxygen for various patient scenarios; however, there is a lack of research on POCs’ efficacy across different respiratory rates specifically among COPD patients. Methods: This bench study was conducted using the IngMar Medical Active Servo Lung 5000 attached to flexible tubing, which simulated an adult patient’s nares. An oxygen analyzer (Maxtec Handi+) was connected to the tubing 6 inches below the nares to estimate the inspired oxygen concentration at the adult trachea. A standard nasal cannula was placed in the simulated patient nares, which was also attached to each POC. The IngMar Lung was programmed to simulate a COPD patient at the following respiratory rates: 15, 20, 30, and 40 breaths/min. POC devices were then compared to an oxygen wall outlet and a standard oxygen concentrator (non-portable CAIRE device). The POCs included the CAIRE FreeStyle Comfort (with Autosat), CAIRE FreeStyle Comfort (without Autosat), Inogen G4, Inogen G5, Phillips SimplyGo Mini, GCE Zen-O lite, Drive Medical iGo2, and Kingon POCs. Descriptive statistics and ANOVAs were computed to determine statistically significant differences between POCS and control devices (wall oxygen and standalone oxygen concentrator). Results: Across most respiratory rate scenarios, the wall oxygen and standalone oxygen concentrator (control group) resulted higher oxygen measurements compared to the POCs. Among the POCs on setting 2, the highest oxygen measurement for 15, 30, and 40 breaths/min was the CAIRE FreeStyle Comfort with AutoSat on sensitivity 5 (24.43%, 24.81%, and 25.0%, respectively). On setting 2, the 20 breaths/min highest oxygen measurement was the Inogen G4 (24.43%). For setting 5, the highest oxygen measurement for 15 breaths/min and 40 breaths/min was the CAIRE FreeStyle Comfort without AutoSat (31.56% and 26.13%, respectively), while the highest oxygen measurement for 20 breaths/min was the Inogen G5 (30.26%), and the highest oxygen measurement for 30 breaths/min was the CAIRE FreeStyle Comfort with AutoSat on sensitivity level 5 (27.89%). Conclusions: The oxygen administered through the wall and the standalone CAIRE oxygen concentrator all delivered higher oxygen levels compared to the POCs (with the exception of the 40 breaths/min scenario on setting 2). However, among the POCs, the CAIRE Freestyle Comfort with AutoSat, CAIRE Freestyle Comfort without AutoSat, and the Inogen G4 all performed the best among the various breathing rate scenarios. %U