TY - JOUR T1 - Influence of Humidification on Comfort During Noninvasive Ventilation With a Helmet JF - Respiratory Care SP - 798 LP - 804 DO - 10.4187/respcare.01735 VL - 58 IS - 5 AU - Kazuyoshi Ueta AU - Toshiji Tomita AU - Akinori Uchiyama AU - Noriyuki Ohta AU - Naoya Iguchi AU - Yukiko Goto AU - Yuji Fujino Y1 - 2013/05/01 UR - http://rc.rcjournal.com/content/58/5/798.abstract N2 - OBJECTIVE: To evaluate optimal humidifier water temperature when using a helmet for noninvasive ventilation. METHODS: Twenty-eight healthy individuals underwent 8 cm H2O CPAP ventilation with FIO2 of 0.21 and 0.5. Each was sequentially tested in the following order: using the helmet without humidification at ambient temperature; with humidification with unheated chamber water; and with humidification with the chamber water at 31°C, 34°C, and 37°C. At each setting, after a 20 min stabilization period, measurements were taken. Comfort level at each setting was evaluated using a visual analog scale rated zero (least comfortable) to 10 (most comfortable). RESULTS: Temperature and relative and absolute humidity inside the helmet increased; however, the comfort scores significantly decreased as the humidification chamber water temperature increased. Regardless of the FIO2, statistically significantly highest comfort scores were obtained when humidification water, with and without active humidification, was at ambient temperature. Unacceptable absolute humidity was obtained only without humidification at room temperature when FIO2 was 0.5. CONCLUSIONS: With the clinical use of a helmet, for patient comfort and mucosal humidification during CPAP, the most desirable conditions are likely to be obtained by humidifying without heating, that is by leaving the water in the humidifier chamber at room temperature. ER -