RT Journal Article SR Electronic T1 Neonatal and Pediatric Manual Hyperinflation: Influence of Oxygen Flow on Ventilation Parameters JF Respiratory Care FD American Association for Respiratory Care SP 2127 OP 2133 DO 10.4187/respcare.02336 VO 58 IS 12 A1 Pricila Mara Novais de Oliveira A1 Armando Augusto Almeida-Junior A1 Celize Cruz Bresciani Almeida A1 Maria Ângela Gonçalves de Oliveira Ribeiro A1 José Dirceu Ribeiro YR 2013 UL http://rc.rcjournal.com/content/58/12/2127.abstract AB BACKGROUND: Although self-inflating bags are widely used for manual hyperinflation, they do not allow ventilation parameters, such as pressure or volume, to be set. We studied the ventilation performance of neonatal and pediatric self-inflating bags. METHODS: We asked 22 physiotherapists to manually hyperinflate 2 lung models (neonatal and pediatric), using self-inflating bags from 3 manufactures (Hudson, Laerdal, and JG Moriya), with flows of 0, 5, 10, and 15 L/min. A pneumotachograph recorded tidal volume (VT), peak inspiratory pressure (PIP), peak inspiratory flow (PIF), peak expiratory flow (PEF), and inspiratory time. RESULTS: The VT, PIP, and inspiratory time delivered by the Hudson, Laerdal, and JG Moriya bags, in both neonatal and pediatric self-inflating bags, were significantly different (P < .001). The PEF and PIF delivered were different only when using the neonatal self-inflating bags (P < .001). The VT, PIP, and PIF delivered with a flow of 0 L/min were lower than with 15 L/min (P < .05) with all the tested bags, in both the neonatal and pediatric sizes. CONCLUSIONS: The performance of the tested neonatal and pediatric bags varied by manufacturer and oxygen flow. There was an increase in VT, PIP, and PIF related to the increase of oxygen flow from 0 L/min to 15 L/min. The neonatal bags showed higher ventilation parameters variation than the pediatric bags.