TY - JOUR T1 - In Vitro Evaluation of Radio-Labeled Aerosol Delivery Via a Variable-Flow Infant CPAP System JF - Respiratory Care SP - 340 LP - 344 DO - 10.4187/respcare.01904 VL - 59 IS - 3 AU - Kimberly D Farney AU - Brandon T Kuehne AU - Laurie A Gibson AU - Leif D Nelin AU - Edward G Shepherd Y1 - 2014/03/01 UR - http://rc.rcjournal.com/content/59/3/340.abstract N2 - BACKGROUND: Nasal CPAP is widely used in neonatal ICUs. Aerosolized medications such as inhaled steroids and β agonists are commonly administered in-line through nasal CPAP, especially to infants with bronchopulmonary dysplasia. We hypothesized that aerosol delivery to the lungs via variable-flow nasal CPAP in an in vitro model would be unreliable, and that the delivery would depend on the position of the aerosol generator within the nasal CPAP circuit. METHODS: We used a system that employed a test lung placed in a plastic jar and subjected to negative pressure. Simulated inspiration effort was measured with a heated-wire anemometer. We used technetium-99m-labeled diethylene triamine penta-acetic acid as our aerosol. The nebulizer was placed either close to the humidifier or close to the nasal prongs in the circuit, and patient effort was simulated with a minute ventilation of 0.4 L/min. RESULTS: Relative aerosol delivery to the infant test lung with the nebulizer close to the humidifier was extremely low (0.3 ± 0.4%), whereas placing the nebulizer close to the nasal prongs resulted in significantly (P < .001) improved delivery (21 ± 11%). Major areas of aerosol deposition with the nebulizer close to the humidifier versus close to the nasal prongs were: nebulizer (10 ± 4% vs 33 ± 13%, P < .001), exhalation limb (9 ± 17% vs 26 ± 30%, P = .23), and generator tubing (21 ± 11% vs 19 ± 20%, P = .86). Placing the nebulizer close to the humidifier resulted in 59 ± 8% of the aerosol being deposited in the inhalation tubing along the heater wire. CONCLUSIONS: Isotope delivery from an aerosol generator placed near the humidifier on variable-flow nasal CPAP was negligible in this in vitro setup; however, such delivery was significantly improved by locating the aerosol generator closer to the nasal CPAP interface. ER -