TY - JOUR T1 - Evaluation of Ventilators for Mouthpiece Ventilation in Neuromuscular Disease JF - Respiratory Care SP - 1329 LP - 1337 DO - 10.4187/respcare.03031 VL - 59 IS - 9 AU - Sonia Khirani AU - Adriana Ramirez AU - Vincent Delord AU - Karl Leroux AU - Frédéric Lofaso AU - Solène Hautot AU - Michel Toussaint AU - David Orlikowski AU - Bruno Louis AU - Brigitte Fauroux Y1 - 2014/09/01 UR - http://rc.rcjournal.com/content/59/9/1329.abstract N2 - BACKGROUND: Daytime mouthpiece ventilation is a useful adjunct to nocturnal noninvasive ventilation (NIV) in patients with neuromuscular disease. The aims of the study were to analyze the practice of mouthpiece ventilation and to evaluate the performance of ventilators for mouthpiece ventilation. METHODS: Practice of mouthpiece ventilation was assessed by a questionnaire, and the performance of 6 home ventilators with mouthpiece ventilation was assessed in a bench test using 24 different conditions per ventilator: 3 mouthpieces, a child and an adult patient profile, and 4 ventilatory modes. RESULTS: Questionnaires were obtained from 30 subjects (mean age 33 ± 11 y) using NIV for 12 ± 7 y. Fifteen subjects used NIV for > 20 h/day, and 11 were totally ventilator-dependent. The subject-reported benefits of mouthpiece ventilation were a reduction in dyspnea (73%) and fatigue (93%) and an improvement in speech (43%) and eating (27%). The bench study showed that none of the ventilators, even those with mouthpiece ventilation software, were able to deliver mouthpiece ventilation without alarms and/or autotriggering in each condition. Alarms and/or ineffective triggering or autotriggering were observed in 135 of the 198 conditions. The occurrence of alarms was more common with a large mouthpiece without a filter compared to a small mouthpiece with a filter (P < .001), but it was not related to the patient profile, the ventilatory mode, or the type of ventilator. CONCLUSIONS: Subjects are satisfied with mouthpiece ventilation. Alarms are common with home ventilators, although less common in those with mouthpiece ventilation software. Improvements in home ventilators are needed to facilitate the expansion of mouthpiece ventilation. ER -