Percentage of inhaled dose delivered through tracheostomy tube via in-line VMN at proximal and distal placement with HHHF in various conditions
InterfaceFlow (L/min)Inhaled dose based on nebulizer placement, % P value
ProximalDistal
Trach adaptor1515.4 ± 1.423.8 ± 1.1.009
309.0 ± .711.5 ± .8.009
453.7 ± .96.8 ± 1.2.009
P value.002.002 
Trach collar1516.4 ± 2.121.3 ± 1.9.009
3013.2 ± 1.18.7 ± .8.009
458.8 ± .95.4 ± .8.009
P value.003.002 
  • Values are presented as mean±SD%; VMN: vibrating mesh nebulizer; HHHF: heated humidified high-flow therapy; Distal: VMN placed near the humidifier; Proximal: VMN placed near the airway.