Transcutaneous arterial carbon dioxide pressure monitoring in critically ill adult patients

  • Technical Note
  • Published:
Intensive Care Medicine Aims and scope Submit manuscript

Abstract

Objective

To evaluate the accuracy of transcutaneous PCO2 (PtcCO2) as a surrogate for arterial PCO2 (PaCO2) in a cohort of adult critically ill patients in a medical intensive care unit (ICU).

Design

Prospective observational study comparing paired measures of transcutaneous and arterial PCO2.

Setting

A 26-bed medical ICU.

Patients

Fifty ICU patients monitored with a SenTec Digital Monitor placed at the ear lobe over prolonged periods.

Results

A total of 189 paired PCO2 measures were obtained. Twenty-one were excluded from analysis, because profound skin vasoconstriction was present (PCO2 bias = −10.8 ± 21.8  mmHg). Finally, 168 were analysed, including 137 obtained during mechanical ventilation and 82 under catecholamine treatment. Body temperature was below 36°C for 27 measurements. Mean duration of monitoring was 17 ± 17 h. The mean difference between PaCO2 and PtcCO2 was −0.2 ± 4.6  mmHg with a tight correlation (R 2 = 0.92, p > 0.01). PCO2 bias did not significantly change among three successive measurements. Changes in PaCO2 and in PtcCO2 between two blood samples were well correlated (R 2 = 0.78, p > 0.01). Variations of more than 8 mmHg in PtcCO2 had 86% sensitivity and 80% specificity to correctly predict similar changes in PaCO2 in the same direction. Catecholamine dose or respiratory support did not affect PtcCO2 accuracy. Hypothermia has only a small effect on accuracy. No complication related to a prolonged use of the sensor was observed

Conclusion

Transcutaneous PCO2 provides a safe and reliable trend-monitoring tool, provided there is no major vasoconstriction.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Hess DR (1998) Capnometry. In: Tobin M (ed) Principles and practice of intensive care monitoring. McGraw-Hill, New York, p 377–400

    Google Scholar 

  2. Simon M, Gottschall R, Gugel M, Fritz H, Mohr S, Klein U (2003) Comparison of transcutaneous and endtidal CO2-monitoring for rigid bronchoscopy during high-frequency jet ventilation. Acta Anaesthesiol Scand 47:861–867

    Article  PubMed  CAS  Google Scholar 

  3. Saura P, Blanch L, Lucangelo U, Fernandez R, Mestre J, Artigas A (1996) Use of capnography to detect hypercapnic episodes during weaning from mechanical ventilation. Intensive Care Med 22:374–381

    Article  PubMed  CAS  Google Scholar 

  4. Lucangelo U, Blanch L (2004) Dead space. Intensive Care Med 30:576–579

    Article  PubMed  CAS  Google Scholar 

  5. Meade MO, Cook DJ (1998) A framework for decision making. In: Tobin M (ed) Principles and practice of intensive care monitoring. McGraw-Hill, New York, p 141–147

    Google Scholar 

  6. Palmisano BW, Severinghaus JW (1990) Transcutaneous PCO2 and PO2: a multicenter study of accuracy. J Clin Monit 6:189–195

    Article  PubMed  CAS  Google Scholar 

  7. Severinghaus JW (1981) A combined transcutaneous PO2-PCO2 electrode with electrochemical HCO3 -stabilization. J Appl Physiol 51:1027–1032

    PubMed  CAS  Google Scholar 

  8. Rodriguez PO, Lellouche F, Aboab J, Brun-Buisson C, Brochard L (2004) Transcutaneous arterial carbon dioxide partial pressure monitoring in ICU patients. Intensive Care Med 30:S168

    Article  Google Scholar 

  9. Tremper KK, Shoemaker WC, Shippy CR, Nolan LS (1981) Transcutaneous PCO2 monitoring on adult patients in the ICU and the operating room. Crit Care Med 9:752–755

    Article  PubMed  CAS  Google Scholar 

  10. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310

    PubMed  CAS  Google Scholar 

  11. Weinstein S, Obuchowski NA, Lieber ML (2005) Clinical evaluation of diagnostic tests. AJR Am J Roentgenol 184:14–19

    PubMed  Google Scholar 

  12. Obuchowski NA (2005) ROC analysis. AJR Am J Roentgenol 184:364–372

    PubMed  Google Scholar 

  13. Rohling R, Biro P (1999) Clinical investigation of a new combined pulse oximetry and carbon dioxide tension sensor in adult anaesthesia. J Clin Monit 15:23–27

    Article  CAS  Google Scholar 

  14. Dullenkopf A, Di Bernardo S, Berger F, Fasnacht M, Gerber A, Weiss M (2003) Evaluation of a new combined SpO2/PtcCO2 sensor in anaesthetized paediatric patients. Paed Anaesth 13:777–784

    Article  Google Scholar 

  15. Eberhard P, Gisiger PA, Gardaz JP, Spahn DR (2002) Combining transcutaneous blood gas measurement and pulse oximetry. Anesth Analg 94:76–80

    Google Scholar 

  16. Janssens JP, Howarth-Frey C, Chevrolet JC, Abajo B, Rochat T (1998) Transcutaneous PCO2 to monitor noninvasive mechanical ventilation in adults. Chest 113:768–773

    Article  PubMed  CAS  Google Scholar 

  17. SenTec (2002) SenTec Digital Monitor – Mode d'emploi. Therwil, Suisse

  18. Severinghaus JW (1982) Transcutaneous blood gas analysis. Resp Care 27:152–159

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Laurent Brochard.

Electronic Supplementary Material

Rights and permissions

Reprints and permissions

About this article

Cite this article

Rodriguez, P., Lellouche, F., Aboab, J. et al. Transcutaneous arterial carbon dioxide pressure monitoring in critically ill adult patients. Intensive Care Med 32, 309–312 (2006). https://doi.org/10.1007/s00134-005-0006-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00134-005-0006-4

Keywords

Navigation