Efficacy of the Airtraq® laryngoscope with a fiberoptic bronchoscope compared with that of Airtraq® alone for tracheal intubation: a manikin study

  • Original Article
  • Published:
Journal of Anesthesia Aims and scope Submit manuscript

Abstract

Purpose

Successful intubation using the Airtraq® requires optimal positioning of the glottis in the middle of the viewfinder. If the glottic view cannot be optimized, some glottic manipulation is essential for the Airtraq-assisted successful intubation. We evaluated the efficacy of the combined use of the Airtraq and a fiberoptic bronchoscope (FOB) compared with that of the Airtraq alone for tracheal intubation in simulated airway scenarios.

Methods

Eight anesthesia providers (four staff and four residents) were enrolled in this study. The participants intubated the trachea of the ALS Simulator manikin in five tongue edema scenarios simulating modified Cormack–Lehane grade 1, 2a, 2b, 3, and 4 views and one cervical immobilization scenario.

Results

No significant difference in the rate of successful intubation was detected between the combined use and the use of Airtraq alone in all scenarios. However, the duration of intubation attempts with the combined use was significantly shorter in difficult laryngoscopy scenarios (Cormack–Lehane grade 2b–4) (P < 0.01) and were significantly longer in easy laryngoscopy scenarios (grade 1 and 2a) (P < 0.05) than those with Airtraq alone. The rate of successful intubation and duration of intubation attempts were similar between the anesthetists and residents in each intubation technique in all scenarios.

Conclusion

The combined use of Airtraq® and a FOB enables rapid intubation in simulated difficult airway scenarios compared with intubation using Airtraq alone, and the speed of intubation performed by anesthetists and residents is similar in all airway scenarios.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Matioc AA. Use of the Airtraq® with a fibreoptic bronchoscope in a difficult intubation outside the operating room. Can J Anesth. 2008;55:561–2.

    Article  PubMed  Google Scholar 

  2. Yentis SM, Lee DJ. Evaluation of an improved scoring system for the grading of direct laryngoscopy. Anaesthesia. 1998;53:1041–4.

    Article  CAS  PubMed  Google Scholar 

  3. Maharaj CH, Costello JF, Higgins BD, Harte BH, Laffey JG. Learning and performance of tracheal intubation by novice personnel: a comparison of the Airtraq and Macintosh laryngoscope. Anaesthesia. 2006;61:671–7.

    Article  CAS  PubMed  Google Scholar 

  4. Ovassapian A. Fiberoptic tracheal intubation in adults. In: Ovassapian A, editor. Fiberoptic endoscopy and the difficult airway. Philadelphia: Lippincott-Raven; 1996. p. 71–103.

    Google Scholar 

  5. Maharaj CH, Higgins BD, Harte BH, Laffey JG. Evaluation of intubation using the Airtraq® or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy: a manikin study. Anaesthesia. 2006;61:469–77.

    Article  CAS  PubMed  Google Scholar 

  6. Maharaj CH, Ni Chonghaile M, Higgins BD, Higgins BD, Harte BH, Laffey JG. Tracheal intubation by inexperienced medical residents using the Airtraq and Macintosh laryngoscopes: a manikin study. Am J Emerg Med. 2006;24:769–74.

    Article  PubMed  Google Scholar 

  7. Asai T, Shingu K. Difficulty in advancing a tracheal tube over a fiberoptic bronchoscope: incidence, causes, and solutions. Br J Anaesth. 2004;92:870–81.

    Article  CAS  PubMed  Google Scholar 

  8. Johnson DM, From AM, Smith RB, From RP, Maktabi MA. Endoscopic study of mechanisms of failure of endotracheal tube advancement into the trachea during awake fiberoptic orotracheal intubation. Anesthesiology. 2005;102:910–4.

    Article  PubMed  Google Scholar 

  9. Will RG, Ironside JW, Zeidler M. A new variant of Creutzfeldt-Jakob disease in the UK. Lancet. 1996;347:921–5.

    Article  CAS  PubMed  Google Scholar 

  10. Lowe PR, Engelhardt T. Prion-related diseases and anaesthesia. Anaesthesia. 2001;56:485.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Kohki Nishikawa.

About this article

Cite this article

Nishikawa, K., Hukuoka, E., Kawagishi, T. et al. Efficacy of the Airtraq® laryngoscope with a fiberoptic bronchoscope compared with that of Airtraq® alone for tracheal intubation: a manikin study. J Anesth 25, 93–97 (2011). https://doi.org/10.1007/s00540-010-1038-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00540-010-1038-7

Keywords

Navigation