Improvement in aerosol delivery with helium-oxygen mixtures during mechanical ventilation

Am J Respir Crit Care Med. 2001 Jan;163(1):109-14. doi: 10.1164/ajrccm.163.1.2003025.

Abstract

In mechanically ventilated patients with airway obstruction, helium-oxygen (He-O2) mixtures reduce airway resistance and improve ventilation, but their influence on aerosol delivery is unknown. Accordingly, we determined the effect of various He-O2 mixtures on albuterol delivery from metered-dose inhalers (MDIs) and jet nebulizers in an in vitro model of mechanical ventilation. Albuterol delivery from a MDI was increased when the ventilator circuit contained 80% helium and 20% oxygen (He-O2 80/20) versus O2: 46.7 +/- 3.3 versus 30.2 +/- 1.3 (SE)% of the nominal dose (p < 0.001)-the difference was mainly due to decreased drug deposition in the spacer chamber, mean 39.2% and 55.2%, respectively (p < 0.001). Nebulizer efficiency at a flow rate of 6 L/min was five times lower with He-O2 80/20 than O2, and the amount of nebulized drug was inversely correlated with gas density (r = 0.94, p < 0.0001). When the nebulizer was operated with O2, greater albuterol delivery was achieved when the ventilator circuit contained He-O2 rather than O2. In summary, He-O2 mixtures in the circuit increased aerosol delivery for both MDIs and nebulizers in the mechanically ventilated model by as much as 50%. In conclusion, at appropriate flow rates and concentrations, He-O2 in the ventilator circuit may improve aerosol delivery in mechanically ventilated patients with severe airway obstruction.

Publication types

  • Evaluation Study
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Aerosols / administration & dosage
  • Helium / administration & dosage*
  • Oxygen / administration & dosage*
  • Respiration, Artificial*

Substances

  • Aerosols
  • Helium
  • Oxygen