Background: The aim of this study was to determine the influence of breathing resistance of heat and moisture exchangers (HMEs) on endotracheal climate and breathing pattern.
Methods: Endotracheal temperature and humidity and tidal volumes were measured in 11 laryngectomized patients with a regularly used HME with "standard" breathing resistance (Provox Normal HME; R-HME), a low breathing-resistance HME (Provox HiFlow HME; L-HME), and without HME.
Results: Both R-HME and L-HME increased end-inspiratory humidity (+5.8 and 4.7 mgH(2)O/L, respectively), decreased end-inspiratory temperature (-1.6 and -1.0 degrees C, respectively), and prolonged the exhalation breath length to approximately 0.5 seconds. The R-HME significantly enlarged tidal volumes (0.07 L; p < .05).
Conclusions: Both HMEs significantly improve tracheal climate. The R-HME has better moistening properties and a small but significant positive effect on tidal volume. Therefore, if the higher resistance is tolerated, the R-HME is the preferred pulmonary rehabilitation device. The L-HME is indicated if lower breathing resistance is required.