Aerosol Generation and Mitigation During Methacholine Bronchoprovocation Testing: Infection Control Implications in the Era of COVID-19

Respir Care. 2021 Dec;66(12):1858-1865. doi: 10.4187/respcare.09236.

Abstract

Background: Methacholine bronchoprovocation or challenge testing (MCT) is commonly performed to assess airway hyper-responsiveness in the setting of suspected asthma. Nebulization is an aerosol-generating procedure, but little is known about the risks of MCT in the context of the ongoing coronavirus disease 2019 (COVID-19) pandemic. We aimed to quantify and characterize aerosol generation during MCT by using different delivery methods and to assess the impact of adding a viral filter.

Methods: Seven healthy subjects performed simulated MCT in a near particle-free laboratory space with 4 different nebulizers and with a dosimeter. Two devices continuously sampled the ambient air during the procedure, which detected ultrafine particles, from 0.02-1 μm, and particles of sizes 0.3, 0.5, 1.0, 2.0, 5.0, and 10 µm, respectively. Particle generation was compared among all the devices, with and without viral filter placement.

Results: Ultrafine-particle generation during simulated MCT was significant across all the devices. Ultrafine-particle (0.02-1 μm) concentrations decreased 77%-91% with the addition of a viral filter and varied significantly between unfiltered (P < .001) and filtered devices (P < .001). Ultrafine-particle generation was lowest when using the dosimeter with filtered Hudson nebulizer (1,258 ± 1,644 particle/mL). Ultrafine-particle concentrations with the filtered nebulizer devices using a compressor were higher than particle concentrations detected when using the dosimeter: Monaghan (3,472 ± 1,794 particles/mL), PARI (4,403 ± 2,948), Hudson (6,320 ± 1,787) and AirLife (9,523 ± 5,098).

Conclusions: The high particle concentrations generated during MCT pose significant infection control concerns during the COVID-19 pandemic. Particle generation during MCT was significantly reduced by using breath-actuated delivery and a viral filter, which offers an effective mitigation strategy.

Keywords: Aerosol; Asthma; COVID-19; Droplet; Infection Control; Methacholine; SARS-CoV-2.