Abstract
BACKGROUND: A high-volume low-pressure endotracheal tube (ETT) cuff forms folds along its contact with the trachea, allowing mucus leakage into the lungs. We developed a thin-walled ETT cuff made of Lycra polyurethane.
METHODS: In vitro, we tested 6 of each of the new prototype Lycra cuff, the Mallinkrodt Hi-Lo ETT (polyvinyl chloride cuff), and the Kimberly-Clark Microcuff ETT (polyurethane cuff), for leakage, in an acrylic mock trachea (inner diameter 20-mm), with a cuff inflation pressure of 20 cm H2O. We poured 15 mL of methylene-blue colored water into the acrylic tube above the cuff and observed for leakage for 24 hours.
RESULTS: The Lycra cuffs had no folds upon inflation in the mock trachea and completely prevented fluid leakage for 24 hours (P < .001 vs the Hi-Lo and the Microcuff). The average leakage past the Hi-Lo was 1,182 ± 1,321 mL/h. The average leakage past the Microcuff was 1.2 ± 0.4 mL/h (P < .001 vs the Hi-Lo).
CONCLUSIONS: Our Lycra cuff provided complete tracheal sealing in vitro.
Introduction
The cuffs of currently available endotracheal tubes (ETTs) are made of polyvinyl chloride (PVC). Such cuffs form an inflatable oblong cuff larger than the average diameter of the trachea, to block mucus flow into the lungs, and to prevent air leakage. Natural rubber latex ETT cuffs are unsuitable for long-term use,1 and the PVC cuff also has limitations. PVC has limited stretch upon inflation, so the cuff diameter needs to be larger than the trachea, and the cuff therefore forms folds along its contact with the trachea,2 which allow passage of pathogen-colonized mucus into the lungs, with a consequent increased risk of pneumonia.
Recently marketed ETT cuffs made of very thin polyurethane (eg, SealGuard, Mallinckrodt/Covidien, Boulder, Colorado) can overcome some of the constraints of previous-generation PVC ETT cuffs. The very thin cuff wall3 forms smaller folds and thus allows less leakage. Mucus leakage past the cuff is a clinically important risk4 that could probably be prevented with a very thin cuff made of a highly distensible elastomer.
Some 20 years ago we developed technology to fabricate thin-walled arterial/venous catheters5 and thin-walled ETTs with Lycra polyurethane, dissolved in N,N-dimethylacetamide.6 Briefly, this entails using a small-gear pump to pump dissolved polyurethane solution onto a steel mandrel rotated by a lathe. The first layer is of polyurethane. After the solvent evaporates we apply a layer of nitinol (nickel/titanium) shape-memory wire for overall strength, then we apply another layer of polyurethane and dry the catheter overnight in a drying oven.
It seemed reasonable to apply this technology (without the wire reinforcement) to the fabrication of a prototype, highly elastic, ultra-thin-walled spherical or cylindrical ETT cuff. The aim of our experiment was to determine, with a static bench model of the trachea, if our Lycra ETT cuff prevents fluid leakage.
Methods
This study was performed at the National Institutes of Health, Bethesda, Maryland, and funded by the intramural research program of the National Heart, Lung, and Blood Institute.
Figure 1 shows the essentials of the manufacturing process. Onto a lathe we mounted a highly polished, straight, Pyrex glass rod of the diameter of the chosen ETT cuff (18 mm). The Lycra solution was l0% solids in N-N-dimethylacetamide, and was filtered before use. It was our goal to fabricate an ultra-thin-walled, cylindrical ETT cuff. For enhanced quality, we applied 2 coats of Lycra, rather than just one thick first coat, to allow for more rapid and uniform drying of the first coat, followed by a second coat of l0% Lycra, thus assuring absence of pinholes.
After 24 hours of drying in an oven at 60°C, the cuff was removed from the glass mandrel and mounted on a standard, commercially available (Mallinckrodt) ETT, from which we had carefully removed the PVC cuff. We secured the Lycra cuff to the ETT with cotton/polyester yarn. We used no adhesives because the Lycra cuff might be weakened by the components of an adhesive. We tested each Lycra cuff for pinholes by inflating the cuff and immersing it under water.
In Vitro Studies
We evaluated the performance of 6 of each of the Lycra ETT cuffs, the (8 mm inner-diameter) Hi-Lo ETT (Mallinckrodt/Covidien, Boulder, Colorado), which has a PVC cuff, and the more recently introduced Microcuff ETT (Kimberly-Clark, Kimberly-Clark Health Care, Roswell, Georgia), which has a polyurethane cuff. An acrylic tube (inner diameter 20 mm)7 was mounted vertically, and after inserting the ETT and inflating the cuff to 20 cm H2O (measured with an aneroid manometer previously tested against a water column) we poured into the acrylic mock trachea l5 mL of methylene-blue colored water, and measured the time needed for all the colored water to leak past the cuff, or, if there was no leak, the ETT was left in place for 24 hours. The cuffs were not lubricated before insertion. The average leakage flow was calculated as the total volume leaked, divided by the time needed for all 15 mL of colored water to leak past the cuff, or 24 hours if not all the colored water leaked past the cuff.
Pressure Transmitted to the Tracheal Wall
When the Lycra cuff is inflated, some of the intracuff pressure is spent to expand the Lycra. The pressure needed to expand the Lycra up to a certain diameter can be calculated from the cuff's pressure/diameter curve, which is linear in the pressure range we explored.6 The diameter of our Lycra cuff is calculated as:
Fig. 2 shows the cuff pressure-diameter relationship. Intracuff pressure was measured with a manometer, and cuff diameter with a caliper.
Statistical Methods
Data are expressed as mean ± standard deviation. We classified each leakage test as with leakage or without leakage. We combined the Lycra cuff tests results with the results from the Microcuff and Hi-Lo on a 2×2 table, and analyzed the values with the Fisher exact test. We compared the absolute leakage with the Wilcoxon test only between the Hi-Lo and the Microcuff, because the leakage with the Lycra cuff showed no standard deviation. All the computations were performed with statistics software (R, R Foundation for Statistical Computing, Vienna, Austria).
Results
Appearance of Inflated Lycra Cuff
The wall thickness of the Lycra cuff was about 10 μm, measured with a dial micrometer. Lycra has a maximum elongation of approximately 500% (Fig. 3), and the cuff surface was smooth and devoid of folds when inflated in the mock trachea. The Lycra cuff touched the wall of the mock trachea, leaving no gap.
In Vitro Leak Test
We observed no leakage of colored water with the Lycra cuff (Fig. 4). All the Hi-Lo cuffs had visible folds that allowed a constant trickle of colored water, with total drainage of all 15 mL in less than 3 min (average flow 1,182 ± 1,321 mL/h). The Microcuff had smaller folds, and the leakage rate was l.2 ± 0.4 mL/h. The number of cuff tests with leakage was zero with the Lycra cuffs, whereas all the Hi-Lo and Microcuff tests had leakage (P < .001). Leakage with the Microcuff was significantly less than that with the Hi-Lo (P < .001).
Transmitted Pressure
The diameter of the Lycra cuff increased approximately 1.4% of the at-rest diameter per cm H2O of intracuff pressure added. An intracuff pressure of approximately 40 cm H2O is needed to increase its diameter 77% from resting position. The at-rest diameter is 17.3 mm, which is slightly less than the glass rod on which the Lycra cuff is manufactured (18 mm), probably owing to drying contraction.
Discussion
Our in vitro results suggest that our low-pressure Lycra cuff can seal the trachea without cuff folds. The commercially available PVC and polyurethane cuffs are almost inelastic and form folds. Absence of folds completely prevents fluid leakage past the cuff, as demonstrated by Young et al, who developed a commercially available thick silicon cuff,8 which, however, requires a high inflation pressure, and by Zanella et al, who developed a prototype double-layer Guayule latex cuff, which is not yet commercially available.9
Our choice of Lycra was based on our previous experience, Lycra's exceptional elongation, and its resistance to deterioration: some Lycra catheters made 30 years ago still look new.5 We developed technology and know-how to fabricate Lycra (and other polymers with high elongation) ETT cuffs that are highly elastic and we believe may improve the care of intubated mechanically ventilated patients.
The present prototype, developed in our laboratory, ensures a perfect seal with a low intracuff pressure if the cuff's resting diameter is close to the diameter of the trachea; if the difference between the cuff's resting diameter and the trachea diameter is greater, higher intracuff pressure is needed because an increased percentage of the intracuff pressure is spent on distending the cuff.
We did not perform a thorough comparison between our cuff and other commercially available ETT cuffs. Our tests were performed at one pressure and one trachea diameter, to test a concept that we believe is fundamental: folds, no matter how small, are invariably associated with possible leakage. The Microcuff design was based on the principle of the traditional high-volume low-pressure cuff: the cuff was meant to be larger than the trachea3; the ultra-thin polyurethane makes the folds very small, which decreased but did not eliminate leakage in our study.9
Subglottic secretions are more viscous than water and therefore flow through narrow folds slower than water, so our choice of water may have biased our results against the Hi-Lo and Microcuff, resulting in a leakage of 1,182 ± 1,321 mL/h with the Hi-Lo, which is not observed clinically. Another possible bias is that a human trachea is neither cylindrical nor rigid, so our results cannot immediately be transferred to clinical practice The lack of water leakage we observed for 24 hours may not reflect clinical performance, since in our setup the mock trachea and ETT remained still, whereas in vivo the cuff moves inside the trachea when the patient's neck bends.10
Though Lycra is highly distensible, it needs some pressure to be inflated and stretched (about 40 cm H2O to obtain a diameter 1.8 times the baseline); part of the intracuff pressure is spent to keep the cuff inflated and part is spent on the tracheal wall, depending on the size of the trachea. When the Lycra cuff is inflated in a trachea with a diameter close to the diameter of the cuff, the transmitted pressure is about 5 cm H2O lower than the intracuff pressure, but with a trachea 5 mm wider the difference between intracuff pressure and transmitted pressure reaches 15 cm H2O.
Young and Blunt proposed a thick cuff devoid of folds,11 which they achieved by inflating the cuff with fixed volume and relying upon the saturation of the shape of the pressure-volume curve beyond a certain pressure value. But that pressure is quite high (> 70 cm H2O). Our Lycra cuff's pressure-volume curve is linear (see Fig. 2) and its slope remains constant, even at low intracuff pressures.
Epidemiologic studies suggest that bacteria are derived from the oropharynx and from the gut, and are responsible for ventilator-associated pneumonia (VAP).12,13 Early-onset VAP seems mostly derived from Gram-positive oral bacteria, whereas late-onset VAP is from multi-resistant Gram-negative bacteria from the gut.12,14 Some bacteria enter the trachea during intubation, but it is reasonable to assume that the bulk of bacteria that cause VAP come from leakage of subglottic secretions past the ETT cuff. Such leakage has also been proven through the finding of pepsin in tracheal secretions.14 It is reasonable to assume that a cuff with no leak would reduce the incidence of VAP. Indirect confirmation of that concept comes from studies on continuous suctioning of subglottic secretions, which found a reduced incidence of VAP, though continuous suctioning is also injurious to tracheal mucosa.15,16 Moreover, recent clinical studies of the Microcuff found a reduced incidence of VAP, which is plausible to associate with reduced cuff leakage.17,18
Conclusions
The substantial risks associated with cuff leakage merit the trial of new technology to prevent cuff leakage and thus improve patient outcomes and lower costs by decreasing the incidence of VAP. In this preliminary bench study we found that our smooth, highly distensible Lycra cuff performed better than Hi-Lo or Microcuff in vitro. Clinical studies of the Lycra cuff are warranted.
Acknowledgments
We thank Angelo Colombo MD for statistical advice.