Abstract
BACKGROUND: Pressure ulcers related to oronasal masks used with noninvasive ventilation (NIV), along with patient discomfort, occur due to improper fit of the mask. We developed a personalized fitting device using a 3-dimensional (3D) scanning solution to prevent the formation of NIV mask-related pressure ulcers. This study aimed to evaluate the effectiveness of the proposed personalized fitting device.
METHODS: We conducted a randomized crossover experimental study of 20 healthy participants to study the use of this personalized fitting device between the face and an NIV mask designed with 3D solutions. The fitting device was not used under the NIV mask for the control. The outcome measures were the presence of blanchable erythema, standardized redness intensity, discomfort level, and contact pressure.
RESULTS: The incidence of blanchable erythema and standardized redness intensity values were significantly lower for subjects who used the fitting device when worn for 30 min (P < .001). The discomfort levels at the forehead, nasal bridge, and both cheeks, as well as leakage, were significantly reduced as well (P = .008, P < .001, P = .001, P = .002, P = .001, P = .02, P < .001, P < .001, P < .001). Contact pressure at the nasal bridge, where pressure ulcers most frequently develop, was significantly decreased with the fitting device (P < .001).
CONCLUSIONS: Personalized fitting devices that incorporate 3D scanning solutions may contribute to the prevention of NIV mask-related pressure ulcers and the reduction of discomfort.
- adverse effects
- critical care
- imaging
- three-dimensional
- intensive care units
- mask fit
- medical device-related pressure ulcers
- noninvasive ventilation
- pressure ulcer
- printing
- three-dimensional
Introduction
Medical device-related pressure ulcers cause pain, loss of function, and infection, resulting in extended hospital stays and increased costs.1 Medical device-related pressure ulcers are recognized as adverse events in health care, but they should be a never event because such pressure ulcers are largely preventable.2,3 In the United States, the development of medical device-related pressure ulcers can lead to significant economic costs.3
Despite the application of best practices to prevent medical device-related pressure ulcers, their prevalence among in-patients is as high as 2.0–6.1% in the ICU setting, and the proportion of medical device-related pressure ulcers to total hospital-acquired pressure ulcers is 20.0–47.1%.4,5 Therefore, more effective strategies to prevent pressure ulcers related to the use of medical devices are urgently needed.
Among these injuries, wounds caused by the improper fit an oronasal mask for noninvasive ventilation (NIV) can drastically reduce the patient's quality of life and sometimes become life-threatening.6–9 The improper fit of the NIV mask also causes patient discomfort.7 An improper fit can lead to pressure ulcers, but sometimes even a properly fitted mask can cause these issues. Blanchable erythema, which signals the potential for pressure ulcer development, can occur quickly, even with short-term use of an improperly fitted NIV mask, especially on the nasal bridge where NIV mask-related pressure ulcers most frequently develop because of thin, delicate, and sensitive skin.6,10–12
To achieve a proper fit, a dressing can be used as a fitting material to eliminate the gap between the patient's face and the mask to redistribute the pressure as well as to minimize leakage volume.9,13,14 In addition, clinicians are educated on the proper application of the NIV mask, especially correct mask placement and angle.13,14 Nevertheless, previous studies have reported that the cumulative incidence of NIV mask-related pressure ulcers is as high as 20.0–60.8%, which is mostly due to difficulty in appropriately fitting the mask.6,11
Raurell-Torreda et al15 associated the occurrence of NIV mask-related pressure ulcers with excessive air leaks. They also reported that clinicians tend to press the NIV mask tightly against the patient's face to decrease air leaks. Experimental studies have determined that a minimum contact pressure to prevent excessive air leaks was 1–2 cm H2O above inspiratory pressure.16,17 However, clinicians do not have a way to confirm the degree of contact pressure required when fitting an NIV mask in the clinical setting.
Thus, appropriate training is needed to properly apply a mask to various face shapes, while minimizing leakage volume and avoiding over-tightening the straps.13,14,18,19 In particular, it may be difficult to fit the mask on emaciated patients, patients using a nasogastric tube, or those who are unable to insert dentures, an issue that can increase the risk of leakage volume.13 Therefore, personalization of the fitting material to fill the gap between the face and mask is necessary to achieve a proper fit.
Objective measurements of the face and mask are required to calculate the gap to design a personalized fitting device for each patient. For this purpose, computed tomography, magnetic resonance imaging, or a 3-dimensional (3D) scanner can be used.20 The use of a 3D scanner was the focus of this study because such a scanner can be applied broadly to patients, from the elderly to young children, because there is no risk of radiation exposure and no need for sedation of the critically ill. Furthermore, the required shape can be easily and immediately obtained with the use of a 3D scanner.10,21,22
There have been some attempts to produce personalized masks or mask cushions with 3D technology to improve patient discomfort due to an NIV nasal mask.23,24 Generally, the use of an NIV nasal mask is preferable for patients with obstructive sleep apnea who need long-term ventilation.25 On the other hand, an NIV oronasal mask is preferred for critically ill patients.26 Manufacturing a personalized NIV oronasal mask is difficult because the facial shapes around both the nose and mouth need to be considered simultaneously. Although such patients are at a high risk for mask-related pressure ulcers, no study has reported the production of a personalized fitting device.27–29
In this study, a personalized fitting device is proposed to fill the gap between the face and the mask to achieve a proper fit using 3D solutions. The objective of this randomized crossover experimental study was to determine whether a personalized fitting device could reduce the incidence of blanchable erythema and discomfort level, as well as redistribute the contact pressure.
QUICK LOOK
Current knowledge
Noninvasive oronasal mask-related pressure ulcers can drastically reduce the patient's quality of life and sometimes become life threatening due to improper fitting of the mask. Despite training for mask fitting, mask-related pressure ulcers still occur.
What this paper contributes to our knowledge
The use of the proposed personalized fitting device with the aid of 3D scanning solutions reduced the incidence of blanchable erythema, standardized redness intensity, discomfort level, and contact pressure on the nasal bridge in healthy participants. A personalized fitting device is one of the options to achieve the proper fit of the mask to prevent mask-related pressure ulcers.
Methods
Workflow for Producing a Personalized Fitting Device
To ensure that the shape of the fitting device was identical to the gap between the face and mask, both surfaces were scanned. To calculate the gap between the face and mask, the mask was fit to the face using 3D software (Artec Studio 10, Artec3D, Luxembourg). Then a mold was produced with a 3D printer (Replicator 2X; MakerBot Industries, Brooklyn, New York). Finally, silicone was cast into the mold to create a personalized fitting device (Fig. 1).
Acquisition of Face and Mask Data
The shapes of the face and NIV oronasal mask (AF811 Oro-Nasal Mask (medium), Philips Respironics, Murrysville, Pennsylvania) were scanned separately using an Eva 3D Scanner (Artec3D). The accuracy and resolution of the scanner were 0.1 mm and 0.5 mm, respectively. When scanning the face, the participant was seated on a chair with the mouth open, as though wearing an NIV mask. The mask was painted using a surfacer (Tamiya, Shizuoka, Japan) to more easily scan the mask.
The fitting device design was based on a clinically ideal fitting so that the mask would float on the face with minimum pressure and leakage volume. A previous study suggested that the positive distance (overlap) between the 3D face and mask may cause NIV mask-related pressure ulcers.12 In light of these findings, pilot studies to achieve ideal mask fitting were performed to determine the best position between the face and mask using 3D software. The negative distance between the nasal bridge and the mask (ie, the space with no contact) was the most important to address to prevent these pressure ulcers. Additionally, to maintain the shape of the fitting device, the device itself should be made of a material harder than that of the NIV mask. However, as the nose is most susceptible to NIV mask-related pressure ulcers,6,11 we focused on the gap between the chin and the mask when creating the fitting device so that it did not come into contact with the area surrounding the nose (Fig. 2). Using 3D software, the 3D face and mask scans were aligned so that the mask floated on facial sites other than the chin.
Personalized Fitting Device and Mold
These 3D data were transformed into solid models using Geomagic Sculpt 3D software (3D Systems, Rock Hill, South Carolina). Torus clay was placed between the face and mask models, then the face and mask models were subtracted from the torus clay using a Boolean operation and unnecessary parts were removed. A mold was designed according to 2 requirements: de-modeling and casting. The personalized fitting device model was subtracted from the mold block. The mold model was divided into top and bottom sections with an injection port created in the top section. A 3D printer (MakerBot Industries) was used to produce the molds. Complete production of these pieces took ∼5 h. Each personalized fitting device was created by casting a silicone (Duplicone; SHOFU, Kyoto, Japan) into the mold. The silicone was composed of 2 types that cured at room temperature in 30 min.
Experimental Methods
Study Design and Participants
A randomized crossover experimental study was conducted in a laboratory at The University of Tokyo. The study protocol was approved by the Ethical Committee of the Graduate School of Medicine, The University of Tokyo (Approval No. 11266). The study participants included 27 healthy individuals age ≥ 20 y recruited between September and November 2016. The exclusion criteria were current infectious respiratory disease, facial skin injury, contact dermatitis in response to medical adhesive tape, photosensitivity, silicone allergy, or intolerance to NIV. After scanning, if the length from the nasal bridge to the inferior border of the lower lip was > 90 mm or the length from the nasal bridge to the tip of the chin was < 100 mm, the participants were excluded because the mask was too big or small (the large and small sizes should be considered for such subjects). We used only the medium-sized NIV mask to focus on the proof-of-concept in this study. In addition, if the gap between the face and mask was too small to create a personalized fitting device, the participant was excluded because a fitting device was not needed.
Study Protocol (Fig. 3)
In a randomized, cross-over experimental study, participants with an NIV mask were randomly assigned to 2 groups. Those in the first group initially used a personalized fitting device between the face and the mask and removed it later, whereas the second group used a personalized fitting device later and not initially. The study was divided into 2 phases: one phase involved the evaluation of the presence of blanchable erythema and associated discomfort, whereas the other phase involved the evaluation of contact pressure. Evaluation of the presence of blanchable erythema and discomfort level with or without the fitting device was performed to leave at least 1 day to avoid any carryover effect. In addition, if both evaluations had to be performed on the same day, contact pressure was evaluated after evaluating erythema and discomfort, considering the effects of the application of sensors to their faces.
An A40 bi-level ventilator and an AF811 oronasal mask (Philips Respironics) were used in this study. The inspiratory positive airway pressure was set at 8 cm H2O, and the expiratory positive airway pressure was set at 4 cm H2O, which were the maximum values considered tolerable to healthy participants. The participants were placed in the supine position on a bed with a head-up angle of 60° in all experiments. The strap tension was measured and recorded using a Stretchable Strain Sensor (Yamaha, Iwata, Japan) attached on the strap.
Blanchable Erythema and Discomfort Level
Determining the presence of blanchable erythema, which predisposes the skin to category I pressure ulcers, was adopted as the main study outcome. Moreover, the discomfort level was subjectively evaluated to confirm whether the fitting device reduced discomfort, pain, oppressive feeling, and leakage at the contact site of the mask.
Before applying the NIV mask, each subject's face was photographed. After fitting the mask, the strap tension was adjusted to ensure that the leakage volume was ∼40 L/min, with or without the fitting device. Afterward, the discomfort levels at the nasal bridge, cheeks, chin, and forehead, as well as discomfort caused by leakage, was measured using a 10-point rating scale in which the higher values indicate greater discomfort. After cessation of ventilation, each participant wore the NIV mask for an additional 30 min, with or without the fitting device, because it was difficult for the participants to tolerate long-term ventilation. After 30 min, the discomfort level was evaluated again, except for discomfort caused by leakage, and photographs were acquired a second time.
The presence of blanchable erythema was evaluated by 2 of the researchers, who were blinded to group allocation. Furthermore, the intensity of redness was evaluated by calculating the maximum Erythema Index between with and without the fitting device around the nasal bridge using ImageJ software (National Institutes of Health, Bethesda MD), as described by Yamamoto et al.30 The redness intensity around the nasal bridge was divided by that around the apex of the nose for standardization because individual complexions differed depending on the time of day. The apex of the nose was selected as the point of standardization because, while the nose was covered by the NIV mask, no pressure was applied to the nose by the mask.
Contact Pressure
We hypothesized that contact pressure at the nasal bridge can be reduced and that the contact pressure values on the face can be distributed evenly with use of the fitting device. To test this hypothesis, 12 sensors (Flexi Force; NITTA Corporation, Tokyo, Japan) were attached to the subjects' faces where the mask or fitting device made contact using double-sided adhesive medical tape (Fig. 4). These sensors were thin enough (0.2 mm) to enable noninvasive measurement. The reliability of contact pressure measurements between the face and mask was previously evaluated.31 The coefficient of variation, measured 3 times in this study, was 13.0% with the fitting device and 12.8% without. After fitting the NIV mask and adjusting the leakage volume to ∼50 L/min, the contact pressure with or without a fitting device was measured 3 times for 1 min each. The leakage volume was increased in consideration of the tendency of the sensors to increase the leakage volume.
The distribution of contact pressure was also evaluated to confirm whether the fitting device could efficiently redistribute the pressure. The distribution of proportion among the 12 sensors was represented as the coefficient of variation of the proportion of contact pressure for each sensor to the total contact pressure, where a smaller value indicates a more even redistribution of contact pressure on the face.
Statistical Analysis
Results were summarized as proportions (%) or medians with interquartile range (IQR). The McNemar test was used to compare the incidence of blanchable erythema with or without use of the fitting device. Continuous variables were compared with or without the fitting device using the Wilcoxon signed-rank test. A P value < .05 was considered statistically significant. All statistical analyses were conducted using SPSS, version 20.0 (IBM, Armonk, New York).
Results
In total, 27 participants were scanned, of whom 7 were excluded because of mismatch in mask size (6 participants) or the gap between the face and mask was too small to create a fitting device (1 participant). Of the 20 participants, 17 (85%) were female, the median age was 34.0 (IQR 29.8–39.0) y, and the median body mass index was 19.7 (IQR 18.9–23.3) kg/m2 (Table 1). None of the study participants dropped out or were lost to follow-up.
All participants displayed blanchable erythema when wearing the mask without a fitting device. On the other hand, 15 of 20 subjects (75%) exhibited no blanchable erythema while wearing a fitting device (P < .001) (Table 2). Standardized redness intensity values for subjects with the fitting device were significantly lower than those for subjects without it (median, 1.0 [IQR 0.8–1.1] vs 1.2 [IQR 1.1–1.3], respectively, P < .001).
The discomfort levels were significantly reduced with the use of the fitting device at all evaluated sites except for the chin, both immediately after mask fitting and 30 min afterward. In addition, the discomfort level of leakage was significantly reduced by the use of the fitting device immediately after mask fitting (P = .001) (Table 3).
The contact pressure on the nasal bridge was significantly decreased with the use of the fitting device (P < .001). In addition, the contact pressure around the chin was significantly increased (P = .03 for Sensor 4, P < .001 for Sensor 5, and P < .001 for Sensor 6). The distribution of contact pressure was significantly lower with the fitting device than without it (P = .01) (Table 4).
The strap tension was significantly higher with the fitting device than without when the leakage volume was maintained at ∼50 L/min (median 105.0 gf [IQR 58.9–138.1] versus 9.4 gf [IQR 0.0–37.9], respectively, P < .001).
Discussion
This is the first study to propose a personalized fitting device created with the aid of 3D solutions to prevent NIV oronasal mask-related pressure ulcers. The incidence of blanchable erythema, standardized redness intensity, discomfort level, and contact pressure on the nasal bridge were reduced with the personalized fitting devices, which were created based on accurate measurements by 3D scanning. Therefore, use of this technology is highly expected to prevent medical device-related pressure ulcers if applied clinically for patients with NIV. Furthermore, this study demonstrated an effective use of 3D scanning and printing technology in health care. Because the major cause of medical device-related pressure ulcers is imperfect size or shape of medical devices,29,32 providing customized care with the use of 3D scanning solutions in the clinical setting could prevent many instances of medical device-related pressure ulcers.
Although the strap tension with a fitting device was significantly higher than that without it, the contact pressure in most participants did not exceed 40 mm Hg, which is the threshold of pressure ulcer formation.33 Dellweg et al17 found that an increased area of contact between the face and the mask reduced the contact pressure during expiration. Personalized fitting devices can increase the contact area by precisely filling the gap between the face and the NIV mask. Notably, the fitting devices dramatically reduced the contact pressure on the nasal bridge. This can be explained by the even distribution of contact pressure, which was confirmed by the distribution of contact pressure. Therefore, the use of the fitting device may reduce the incidence of mask-related pressure ulcers.
Furthermore, the application of a fitting device might be more effective than a dressing in preventing pressure ulcers. A previous study14 evaluated the effectiveness of various dressings for the prevention of NIV mask-related pressure ulcers by checking the skin condition of ICU subjects every 30 min. In the control group, which received no dressing application, 96.7% developed medical device–related pressure ulcers, while 53.3% and 40.0% of subjects who received film or hydrocolloid dressing, respectively, developed medical device-related pressure ulcers.14 As indicated by our results, the use of this fitting device considerably reduced the incidence of blanchable erythema, implying potential effectiveness to prevent medical device-related pressure ulcers. Although the incidence of such pressure ulcers was not evaluated, we believe that the use of the proposed fitting device could reduce the incidence because its use is not dependent on the clinician's skill.
In this study, the use of this fitting device resulted in blanchable erythema in 5 subjects even when the contact pressure was reduced. Use of a fitting device requires greater strap tension than a NIV mask only to maintain the fitting device on the face. The weight of the fitting device may have been the cause of the high-contact pressure around the chin as well as the shear stress on the nasal bridge, which is a known risk factor for NIV mask-related pressure ulcers.6 Therefore, we plan to use lighter materials to create the fitting devices in future studies.
This study has several limitations. First, the subjects were not blinded to intervention assignment, which might have affected the reporting of discomfort level. Second, our study was limited to Asian participants with a limited variety of facial characteristics. Third, the number of subjects was small and the subjects were healthy individuals, so the results may differ in patients who differ from the study population in terms of age, nutritional status, medication use, presence of comorbidities, and level of consciousness.13 Fourth, the short duration of the fitting device use did not allow an adequate assessment of the effects of the fitting device after 4–6 h under a sustained load, which would be an appropriate time frame for NIV mask-related pressure ulcers.34 Fifth, we could not consider other prevention strategies such as mask rotation because the duration of NIV mask application was limited.
Despite these limitations, we have proposed a promising method to prevent NIV mask-related pressure ulcers using 3D scanning solutions. The capacity of the proposed fitting device to prevent NIV mask-related pressure ulcers should be further investigated.
Conclusion
The use of the proposed personalized fitting device created with the aid of 3D scanning solutions reduced the incidence of blanchable erythema, standardized redness intensity, discomfort level, and contact pressure on the nasal bridge in healthy participants. The results of this study imply that the use of personalized fitting devices could prevent NIV mask-related pressure ulcers. Further studies are required to confirm the effectiveness of this device in the clinical setting.