Abstract
BACKGROUND: Stress index provides a noninvasive approach to detect injurious ventilation patterns and to personalize ventilator settings. Obtaining the stress index (SI), however, requires quantitatively analyzing the shape of pressure-time curve with dedicated instruments or a specific ventilator, which may encumber its clinical implementation. We hypothesized that the SI could be qualitatively determined through a visual inspection of ventilator waveforms.
METHODS: Thirty-six adult subjects undergoing volume controlled ventilation without spontaneous breathing were enrolled. For each subject, 2 trained clinicians visually inspected the pressure-time curve directly from the ventilator screen. They then qualitatively categorized the shape of pressure-time curve as linear, a downward concavity, or an upward concavity at the bedside. We simultaneously recorded airway pressure and flow signals using a dedicated instrument. A quantitative off-line analysis was performed to calculate the SI using specific research software. This quantitative analysis of the SI served as the reference method for classifying the shape of the pressure-time curve (ie, linear, a downward concavity, or an upward concavity). We compared the SI categorized by visual inspection with that by the reference.
RESULTS: We obtained 200 SI assessments of pressure-time curves, among which 125 (63%) were linear, 55 (27%) were a downward concavity, and 20 (10%) were an upward concavity as determined by the reference method. The overall accuracy of visual inspection and weighted kappa statistic (95% CI) was 93% (88–96%) and 0.88 (0.82–0.94), respectively. The sensitivity and specificity to distinguish a downward concavity from a linear shape were 91% and 98%, respectively. The respective sensitivity and specificity to distinguish an upward concavity from a linear shape were 95% and 95%.
CONCLUSIONS: Visual inspection of the pressure-time curve on the ventilator screen is a simple and reliable approach to assess SI at the bedside. This simplification may facilitate the implementation of SI in clinical practice to personalize mechanical ventilation. (ClinicalTrials.gov registration NCT03096106.)
- mechanical ventilation
- respiratory mechanics
- stress index
- ventilator-induced lung injury
- personalized medicine
Introduction
Mechanical ventilation, while providing necessary ventilatory support and gas exchange, may induce lung injury.1 Alveolar overdistention and tidal recruitment are 2 main mechanisms of ventilator-induced lung injury.2 To limit the risk of overdistention and tidal recruitment, tidal volume (VT) and PEEP are adjusted.3,4 However, optimizing VT and PEEP for each individual patient is challenging, partly due to the lack of an easy-to-use method to evaluate overdistention and tidal recruitment at the bedside.5
Assessing respiratory mechanics is essential to individualize ventilator settings to minimize ventilator-induced lung injury.6 Analysis of the dynamic airway pressure-time (Paw-t) curve profile is a noninvasive, real-time method for respiratory monitoring at the bedside. Since it was first introduced in 2000,7 analysis of Paw-t curve profiles has been used to identify tidal recruitment and overdistention in both animal experiments and clinical studies.7–16 During volume controlled ventilation, the shape of the Paw-t curve within the constant inspiratory flow section is related to the change of respiratory system elastance. Paw-t curves that are linear, have a downward concavity, or have an upward concavity indicate that the elastance is constant, decreasing, or increasing during tidal inspiration, respectively.17,18 The stress index (SI), which is a dimensionless coefficient derived from curve fitting, can quantitatively describe the shape of the Paw-t curve. However, it requires dedicated data acquisition instruments and analytic software. To our knowledge, there is only one type of ventilator that incorporates the measurement of SI (Servo-i, Maquet, Solna, Sweden) which requires a pre-installed software module. The requirements for specific hardware and software may limit the implementation and use of SI in clinical practice. Some academic centers in the United States have incorporated visual inspection of SI at the bedside. To our knowledge, however, a systematic visual method has not been reported and the accuracy of such a simplified approach has never been validated.
Because different SI values correspond to different Paw-t curve shapes, we speculated that SI could be qualitatively assessed by visually inspecting the waveforms on a ventilator screen. A qualitative assessment would be sufficient to identify injurious ventilation patterns such as overdistention and tidal recruitment. Our aim was to provide a simple and reliable method for personalizing ventilator settings at the bedside. We tested the accuracy and the inter-observer reliability of the visual inspection method in this study.
QUICK LOOK
Current knowledge
Mechanical ventilation provides ventilatory and gas exchange support but may also induce lung injury. The stress index provides a noninvasive approach to detect injurious ventilation patterns and to personalize ventilator settings. However, obtaining the stress index requires quantitatively analyzing the shape of airway pressure-time curve with specific analytical instruments, which may limit its clinical implementation.
What this paper contributes to our knowledge
Our study describes a valid and reliable method for visual inspection of stress index. The results support the suggestion that this simple approach can be utilized quickly and accurately at the bedside to provide optimal, personalized ventilation.
Methods
Study Ethics and Population
This prospective observational study was conducted in the neurosurgical ICU of the Beijing Tiantan Hospital, Capital Medical University, Beijing, China. The protocol was approved by the local institutional review board (KY 2017-016-02) and was registered at ClinicalTrials.gov (NCT03096106). Written informed consent was obtained from each subject's appropriate substitute decision maker because the subject was unconsciousness at enrollment.
Patients were screened daily at 10:00 am for study eligibility. The same subject was allowed to test repeatedly on different days. Inclusion criteria were age 18–80 y, intubated and ventilated with volume controlled ventilation using constant flow, and deep sedation and absence of spontaneous breathing (ie, no triggering during tidal breaths and no inspiratory effort during a 5-s end-expiratory hold). Exclusion criteria were evidence of active air leak from the lung, including bronchopleural fistula, pneumothorax, pneumomediastinum, or existing chest tube; chest wall and/or abdominal injuries; and evidence suggesting reduced chest wall compliance, such as existing large pleural effusion, thoracic trauma, and or intra-abdominal hypertension (ie, intra-abdominal pressure >20 mm Hg).
Paw-t Curve Analysis and SI Classification With Dedicated Software
After enrollment, a heated Fleisch pneumotachograph (Vitalograph, Lenexa, Kansas) was placed between the Y-piece of the ventilator circuit and the endotracheal tube to measure flow. A KT 100D-2 pressure transducer (KleisTEK, Bari, Italy) was connected proximally to the endotracheal tube to measure the Paw. Signals were continuously displayed and saved (ICU-Lab 2.5 software package, KleisTEK) on a laptop computer for further analysis, at a sample rate of 200 Hz.
The Paw-t curve profile analysis was performed using the standard method (see the supplementary materials at http://www.rcjournal.com).7,10,13,15 The constant flow segment was identified as the portion of the waveform where the fluctuation was within ± 3% of the steady value. To eliminate the influence of on-flow and off-flow transients, the segment was further narrowed for 50 ms after the beginning and before the end of the constant flow portion (see the supplementary materials). If a constant segment could not be found or the length of selected segment was less than one third of the entire inspiratory phase, the data were discarded but documented. After confirming the constant flow segment, SI was automatically calculated by fitting the Paw-t curve within the corresponding segment (see the supplementary materials).7 The shape of each Paw-t curve was classified into 3 categories according to the SI value: a linear shape with SI = 0.9–1.1, a downward concavity shape with SI < 0.9, and an upward concavity shape with SI > 1.1 (see the supplementary materials).18
Visual Classification of SI and Training
A standardized method was developed for visual SI estimation by inspecting the waveforms on the ventilator screen (Fig. 1). The midpoint on the constant inspiratory flow was identified, and then the corresponding point on the Paw-t waveform was confirmed. We then put a ruler on the ventilator screen, with the straight edge passing through the point as the reference of tangent line of the Paw-t waveform (Fig. 1A). The relationship of this ruler and the Paw-t waveform was visually inspected and classified as a linear shape if the Paw-t waveform almost coincident with the ruler, indicating an SI = 0.9–1.1 (Fig. 1B); as a downward concavity if the 2 sides of the Paw-t waveform both deviated downward from the ruler and the curve was judged as a downward concavity shape, indicating an SI < 0.9 (Fig. 1C); or as an upward concavity if the 2 sides of the Paw-t waveform both deviated upward from the ruler and the curve was judged as an upward concavity shape, indicating an SI > 1.1 (Fig. 1D).
Four critical care physicians were trained to perform the visual classification. For the pre-test training, 20 subjects' data with ventilator screen photographs and off-line SI analysis were collected. One expert conducted 2 rounds of group training. The observers were asked to discriminate the shape of the Paw-t waveform as linear, a downward concavity, or an upward concavity with a training set of ventilator screen photographs. During group training, the observers reported that it was difficult to distinguish the upward concavity shape from the linear shape. A special instruction was emphasized that the upward concavity shape should only be confirmed when both sides of the Paw-t waveform deviated from the tangent line (ruler) (Fig. 1D). The training manual is available in the supplementary materials.
On-Site Visual SI Classification and Data Collection
After enrollment, we collected the following information: ventilator brands and clinical settings, and subjects' baseline characteristics and physiological data. Three ventilator brands were used in our unit: Servo-i (Maquet), Avea (CareFusion, Mettawa, Illinois), and Puritan Bennett 840 (Covidien, Tyco Healthcare International Trading, China). The investigators were not involved in any clinical decisions. Ventilator settings were not changed during the on-site test. The data acquisition system was set up, and flow and Paw tracings were closely observed to confirm that the subject did not have any spontaneous breathing efforts. A 5-s end-expiratory occlusion was performed, and the ventilator screen displaying flow-time and Paw-t waveforms was frozen to ensure that the screen contained the entire breath just before the occlusion, which was used as the sample breath for visual SI classification and subsequent off-line software analysis. The test was terminated if the subject exhibited spontaneous inspiratory effort during occlusion. In each on-site assessment, 2 observers were randomly chosen from the 4 trained physicians to independently perform visual SI classification at the bedside using our method. The trained observers were instructed to classify the sample breath as having a linear, a downward concavity, or an upward concavity shape, and the results were documented.
Flow and Paw data were collected for off-line analysis. SI was measured by an experienced investigator, who was blinded to the result of visual classification. The same sample breath used for the visual classification was used for the quantitative analysis, (ie, the breath just before the end-expiratory occlusion). Therefore, for each tested breath, on-site visual SI classifications by the 2 observers were accompanied by an off-line software fitting result. In the first 50 off-line analyses, we also measured the averaged SI value during the 2-min period before the sample breath.
Sample Size
Sample size was calculated using Power Analysis and Sample Size (PASS 11, NCSS, Kaysville, Utah). The primary end point was the agreement between the visual SI classification and the reference. Our pre-training pilot data showed that the prevalence of linear, downward concavity, and upward concavity Paw-t curves using off-line software fitting was 75%, 15%, and 10%, respectively. Thus, 192 assessments were required to reject the null hypothesis that the kappa statistic was equal to 0.4 for a 2-tailed test with the type-1 error probability of 0.05 and type-2 error probability of 0.2. To compensate for unqualified data, 220 assessments were collected in the formal test.
Statistical Analysis
SI assessed by the off-line software fitting was used as the reference standard. Agreement between the visual SI classification and the reference standard, as well as inter-observer reliability, was evaluated using a weighted kappa statistic and 95% CI. A weighted kappa statistic > 0.80 is considered as excellent agreement.19 For the first and the second on-site observer, the result with the higher weighted kappa statistic was used for further analysis of diagnostic accuracy. The overall accuracy (95% CI) for visual SI classification was reported. The chi-square test was used to compare the accuracy among the 3 SI classifications and among different ventilator brands. With respect to the reference, sensitivity, specificity, positive predictive value, and negative predictive value for visual SI classification to distinguish downward concavity or upward concavity, the deviation of the Paw-t waveform shape from a linear shape was calculated.20 In particular, the diagnostic accuracy of each and combined upward concavity and downward concavity classification were assessed against the classification of linear shape. Statistical analyses were conducted using SPSS V.20.0 (SPSS, Chicago, Illinois). A P < .05 was considered statistically significant.
Results
We enrolled 36 subjects and collected 220 SI assessments (Table 1); 10 assessments were excluded due to spontaneous breathing effort after enrollment, and another 10 assessments were excluded as the length of the constant flow segment was less than one third of the inspiratory phase (see the supplementary materials). We eventually analyzed 200 qualified assessments obtained from Servo-i (41%), Avea (31%), and Puritan Bennett 840 ventilator (29%). The median (interquartile range) of tidal volume (VT) per predicted body weight and PEEP were 7.2 (6.5–8.6) mL/kg and 8 (6−10) cm H2O, respectively. Using off-line software analysis (reference standard), the values of SI were classified as having a linear, a downward concavity, or an upward concavity shape in 125 (63%), 55 (27%), and 20 (10%) assessments, respectively (Table 2). In the first 50 off-line analyses, the Paw-t curve shape assessed by the SI value of the single sampled breath was identical to that determined with the 2-min mean value in each assessment (see the supplementary materials).
Agreement between visual SI classification by the first and the second observer and the reference standard was substantial, with weighted kappa (95% CI) being 0.86 (0.80–0.92) and 0.88 (0.82–0.94), respectively. Inter-observer reliability was near perfect, with weighted kappa (95% CI) 0.96 (0.92–0.99). Thus, results from the second observer were used for further analysis.
The overall accuracy (95% CI) for visual SI classification was 93% (88–96%). There was no significant difference in accuracy among reference SI classifications (P = .20) or among different ventilator brands (P = .68) (see the supplementary materials).
The contingency table for diagnostic accuracy analysis is shown in the supplementary materials. The sensitivity, specificity, positive predictive value, and negative predictive value of visually detecting a downward concavity shape as compared to a linear shape were 91%, 98%, 96%, and 96%; these respective values for visually detecting an upward concavity shape as compared to a linear shape were 95%, 95%, 76%, and 99% (Table 3). When reference classifications of a downward concavity and an upward concavity were combined as the positive result, the sensitivity and specificity for visual SI classification were 92% and 94%, respectively.
Discussion
In this study, we introduced a method for estimating SI at the bedside by visually inspecting the waveforms on the ventilator screen. Our main finding is that visual SI classification shows an excellent agreement with the reference method, which used off-line software fitting. The visual inspection method can be used to simply, accurately, and reliably distinguish the Paw-t curve shapes, which can help to identify potentially injurious ventilator settings in a timely manner at the bedside.
The Paw-t curve profile analysis can be used to guide VT and PEEP settings during mechanical ventilation.17,18 Experiments in a porcine acute lung injury model showed that an SI < 0.9 indicated tidal recruitment and an SI > 1.1 indicated overdistention.10,11 Moreover, histologic and inflammatory markers of injurious ventilation were minimized when the ventilator was set to maintain SI in the range of 0.9–1.1.7–9 In subjects with ARDS, titrating PEEP to obtain SI of 0.9–1.1 attenuated alveolar overdistention, decreased lung elastance, and increased recruitment volume and oxygenation.13,14 Based on these results, we selected 0.9 and 1.1 as the thresholds for SI classification in our study.
Several conditions are required for accurate and reliable Paw-t curve analysis.17,18 We excluded patients with symptoms of impaired chest wall elastance, which may influence the SI interpretation.21 A standard algorithm was used for off-line software fitting at a high sample rate (200 Hz). A stable constant flow inspiration is critical for the SI determination.17,18 One of the disadvantages of SI is that it cannot be measured under spontaneous breathing. Thus, we excluded the presence of spontaneous effort by performing an end-expiratory occlusion before each assessment of SI. In previous reports, flow and pressure data were recorded for at least 2 min to obtain a mean SI value.7,10,15 However, only a single breath can be visually inspected on the ventilator screen. In a pre-specified analysis, we found identical SI classifications determined with a single sample breath and with the mean SI value of 2-min recorded breaths, which supports the ability to analyze only a single breath. Using Paw-t curve to assess the change in compliance assumes that the resistive pressure remains unchanged at a constant flow during inflation. This assumption can be challenged in a few circumstances when airway resistance is varied during inflation, such as airway closure and reopening. We therefore decided to test the accuracy of SI assessment using clinical ventilator settings with various flow and PEEP levels. We did not set a low flow (eg, <10 L/min) to assess SI in our study. Although the low flow may minimize the resistive pressure and might accentuate the shape for analysis, there are several practical issues and a theoretic issue that preclude us from using it. First, air leak, especially micro leaks around the endotracheal cuff and circuit connections, are extremely difficult to eliminate completely in clinical practice. Using a low flow can significantly amplify the air leak because the inspiratory time is much longer. In this case, any drop on the Paw during inflation can greatly influence the shape of Paw-t curve, leading to potential misinterpretations. Second, flow sensors on ventilators are usually less accurate and less sensitive to a low flow. Third, we intentionally kept the assessment as simple as possible (ie, without the need to change the ventilator settings for clinicians). Finally, the alveolar recruitment process may depend on the different time constant in different lung regions, and a tidal recruitment that occurs in a low-flow inflation does not necessarily mean it would occur in a regular flow. Similarly, overdistention during a regular flow inflation does not necessarily mean it would occur in a low-flow inflation.
The Paw-t curve may change from a linear shape to an upward concavity or to a downward concavity depending on the amount of VT delivered.10 We did not alter the clinical settings of the ventilator during the study. In addition, the delivered VT was higher in the upward concavity category than in the downward concavity and linear categories (Table 2). During mechanical ventilation, a patient's respiratory mechanics may change dynamically over time. Thus, repeated adjustments of ventilator settings are needed to minimize the risks and prevent ventilator-induced lung injury. Simple bedside monitoring that provides real-time answers is required for a method to be useful to clinicians. Our results confirmed the hypothesis that SI can be assessed by visually inspecting the waveforms on the ventilator screen. Visual SI classification yielded excellent agreement with off-line software determination, excellent inter-observer reliability, and high overall accuracy independent of SI classifications and ventilator brands. There are 3 core components in our visual SI classification method (Fig. 1). First, confirmation of the midpoint on a square flow waveform ensures to the greatest extent possible that the period with steady-state flow is selected for visual inspection. Second, placing a ruler on the ventilator screen to mark the tangent line is easy to do at the bedside. Third, visually comparing the shape of the pressure waveform with the tangent line is the key step. During group training, we noticed that mistakes were more likely to occur in distinguishing the upward concavity shape from the linear shape. The observers were specifically instructed that the upward concavity shape should only be confirmed when both sides of the Paw-t waveform deviate from the tangent line (Fig. 1D). Nonetheless, although acceptable sensitivity and specificity were obtained, a low positive predictive value (Table 3) was found in the use of visual classification to detect an upward concavity shape against a linear shape during the formal on-site test, which suggested that training should be emphasized on the identification of an upward concavity shape.
For Paw-t curve profile analysis, the clinical condition of interest lies in its ability to quickly and accurately detect the shape that is consistent with “non-protective” ventilator settings.15 Using software-derived SI classification as the reference, excellent diagnostic accuracy was found for this visual inspection method in distinguishing the downward or upward concavity shape from the linear shape. A single visual classification typically took <30 s, which may facilitate the process of monitoring and ventilator settings titration at the bedside. Further studies are needed to investigate the clinical feasibility of the visual inspection method in the adjustment of ventilator settings.
There are several limitations to our study. First, the inspiratory flow was set at approximately 40 L/min during our routine clinical practice. We did not test the visual inspection method at higher flows. Second, the identification of a stable flow segment is critical in Paw-t curve analysis.17,18 An unqualified constant flow segment occurred in nearly 5% (10 of 220 assessments) in our off-line fitting, but this situation was not observed during on-site visual inspection. Third, whereas a sigmoid shape of the Paw-t curve has been found in animal models of ARDS,9,10 we did not find an obviously sigmoid-shaped curve during visual inspection or with off-line nonlinear fitting, possibly due to the low VT that we used. Alternatively, this difference may result from the degree of the lung inhomogeneity. We speculate that the lungs were relatively homogenous in our subjects, most of whom were surgical and neurological patients. In patients with ARDS who have more inhomogeneous lungs, it is possible that a sigmoid-shaped Paw-t curve may be observed, leading to difficulties for analysis using visual inspection. Finally, the distribution of SI (linear, downward concavity, or upward concavity) may differ in other types of subjects, such as those with ARDS.17,18
Conclusion
This study demonstrates that SI can be accurately and reliably assessed by visually inspecting waveforms on the ventilator screen. This simplified method may help care providers use SI in clinical practice or in large trials for personalized mechanical ventilation.
Acknowledgments
We thank Mr Dawei Wu for his excellent drawing of the ventilator screen in Figure 1.
Footnotes
- Correspondence: Jian-Xin Zhou MD, Department of Critical Care Medicine, Beijing Tiantan Hospital, Capital Medical University, No 6, Tiantan Xili, Dongcheng district, 100050, Beijing, China. E-mail: zhoujx.cn{at}icloud.com.
This study was supported by grants from the Beijing Municipal Administration of Hospital (ZYLX201502, DFL20150502). The sponsor had no role in the study design, data collection, data analysis, data interpretation, or writing of the report. The authors have disclosed no conflicts of interest.
Supplementary material related to this paper is available at http://www.rcjournal.com.
See the Related Editorial on Page 1194
- Copyright © 2018 by Daedalus Enterprises