Abstract
BACKGROUND: During continuous flow CPAP for noninvasive respiratory support, a high flow (eg, 60–90 L/min) of gas with FIO2 titratable up to 1.0 is provided within a helmet or face mask, while a PEEP valve maintains the set pressure. A large amount of oxygen is wasted, whereas only a minimal amount is consumed. We describe a recirculation circuit designed to reuse the exhaust gas and save oxygen.
METHODS: A standard Venturi-based continuous flow system delivering a flow of ≥ 60 L/min to a helmet, and a modified system designed to recirculate oxygen were tested on the bench during simulated active breathing. The proposed system recirculates the oxygen-enriched gas escaping from the PEEP valve; after CO2 removal, the gas is entrained by the Venturi flow generator and redirected to the helmet. We compared oxygen consumption, pneumatic performance, and gas conditioning of the standard and the recirculation systems.
RESULTS: The recirculation system reduced the oxygen consumption up to 80% as compared to the standard system. Oxygen sparing increased with increasing FIO2 and total flow delivered to the helmet. Exhaled CO2 was efficiently removed by a single soda lime canister for about 10 h. Pressure swings during a respiratory cycle slightly worsened when using the recirculation system as compared to the standard one (from 3.5–4.4 cm H2O). At FIO2 1.0, humidity was 6.1 mg/L and 17.3 mg/L with the standard and the recirculation system, respectively.
CONCLUSIONS: The recirculation system allowed a 80% reduction of oxygen consumption during simulated helmet CPAP therapy, whereas CO2 removal was effective for > 10 h. Recirculation minimally affected pneumatic performance of the CPAP continuous flow system, while improving gas conditioning as compared to the standard system.
- compressed oxygen recirculation
- continuous flow CPAP
- head helmet
- humidity and gas water content
- acute hypoxic respiratory failure
Footnotes
- Correspondence: Giacomo Bellani MD PhD, University of Milan-Bicocca, Department of Medicine and Surgery, Via Cadore 48, Monza (MB), Italy; E-mail: giacomo.bellani1{at}unimib.it
See the Related Editorial on Page 286
Drs Coppadoro and Bellani disclose a relationship with Flow-Meter and are coinventors of a pending patent relative to the presented device. Mr Paratico is employed by Flow-Meter. Dr Bellani discloses relationships with Dräger Medical, Getinge, and Siaretron.
- Copyright © 2023 by Daedalus Enterprises
Pay Per Article - You may access this article (from the computer you are currently using) for 1 day for US$30.00
Regain Access - You can regain access to a recent Pay per Article purchase if your access period has not yet expired.