Abstract
BACKGROUND: Accurate measurement of carbon dioxide elimination (V̇CO2) and oxygen consumption (V̇O2) at the bedside may help titrate nutritional and respiratory support in mechanically ventilated patients. Continuous V̇CO2 monitoring is now available with many ventilators. However, because normative data are sparsely available in the literature, we aimed to describe the range of V̇CO2 and V̇O2 values observed in mechanically ventilated children. We also aimed to examine the characteristics of V̇CO2 values that are associated with standard steady state (5-min period when V̇CO2 and V̇O2 variability are < 10%).
METHODS: Mechanically ventilated patients who underwent indirect calorimetry testing were eligible for inclusion, and subjects who achieved standard steady state were included. Normalized V̇CO2 and V̇O2 values (mL/kg/min) were modeled against subject height, and correlation coefficients were computed to quantify the goodness of fit. A steady-state definition using only V̇CO2 was developed (V̇CO2 variability of < 5% for a 5-min period) and tested against standard steady state using sensitivity and specificity.
RESULTS: Steady-state data from 87 indirect calorimetry tests (in 70 subjects) were included. For age groups < 0.5, 0.5–8, and > 8 y, the mean V̇CO2 values were 7.6, 5.8, and 3.5 mL/kg/min. Normalized V̇CO2 and V̇O2 values were inversely related to subject height and age. The relationships between normalized gas exchange values and height were demonstrated by the models: V̇CO2 = 115 × (height in cm)−0.71 (R = 0.61, P < .001) and V̇O2 = 130 × (height in cm)−0.72 (R = 0.61, P < .001). Steady-state V̇CO2 predicted standard steady state (sensitivity of 0.84, specificity of 1.0, P < .01).
CONCLUSIONS: V̇CO2 and V̇O2 measurements correlated with subject height and age. Smaller and younger subjects produced larger amounts of CO2 and consumed more O2 per unit of body weight. The use of a 5-min period when V̇CO2 varied by < 5% predicted standard steady state. Our observations may facilitate greater utility of V̇CO2 at the bedside in the pediatric ICU and thereby extend the benefits of metabolic monitoring to a larger group of patients.
- carbon dioxide elimination
- oxygen consumption
- mechanical ventilation
- children
- pediatric
- V̇CO2
- volumetric capnography
- steady state
Introduction
The integration of CO2 and flow measurements yields carbon dioxide elimination (V̇CO2) measurements that can be a useful metric to detect changes in alveolar ventilation, pulmonary perfusion, CO2 production, and ventilation optimization.1–5 Similarly, volumetric oxygen consumption (V̇O2) can be extrapolated using oxygen and flow to detect the volume of oxygen consumed by the body in a given time period. The combination of V̇CO2 and V̇O2 is used to calculate energy expenditure and plays an important role in assessing and optimizing nutrition, particularly during critical illness.1
Bedside measurements of V̇CO2 and/or V̇O2 in mechanically ventilated patients can be obtained using a standalone bedside monitor, a device integrated into the ventilator, or an indirect calorimeter.6,7 The ability to detect abnormally high or low values could enhance bedside decision support. Hence, reference ranges for these gas exchange measurements are desirable in the critically ill population but are currently not available. We describe observed gas exchange data (both V̇CO2 and V̇O2) in a cohort of critically ill, mechanically ventilated subjects admitted to the pediatric ICU (PICU).
To ensure that the values for V̇CO2 and V̇O2 measured at the respiratory interface reflect cellular CO2 production and O2 utilization by the body, a period of steady state must be identified.8–10 Recently, V̇CO2 was shown to have an expanding role at the bedside and may be sufficient to predict energy expenditure in the majority of mechanically ventilated children.11 However, methods by which a steady-state interval can be computed on stand-alone V̇CO2 monitors to ensure data quality have not been demonstrated in the literature and are needed. The current standard definition of a steady-state period requires 5 consecutive min during which V̇CO2 and V̇O2 vary by ≤ 10%.9 Because many devices incorporate V̇CO2 measurements independently of V̇O2 (such as those incorporated into ventilators), a steady-state definition for V̇CO2 is required when interpreting bedside measurements. Recently, V̇CO2 values were shown to provide useful metabolic data and reasonable estimates of energy expenditure in mechanically ventilated children,11 and methods to determine steady-state conditions with V̇CO2 are desirable. Therefore, we sought to identify a definition for steady state that includes parameters measured by stand-alone V̇CO2 monitors or ventilators and that is in agreement with steady-state V̇CO2 values using the standard definition. This would allow reliable interpretation of the V̇CO2 values at the bedside.
QUICK LOOK
Current knowledge
The measurement of carbon dioxide production (V̇CO2) and oxygen consumption (V̇O2) with instruments integral to a ventilator or with an independent device can provide data regarding nutritional requirements and ventilatory efficiency. The presence of leaks in pediatric patients complicates these measurements and can introduce important errors. Establishment of a steady state in this population can be difficult.
What this paper contributes to our knowledge
V̇O2 and V̇CO2 measurements correlated with subject height and age. Smaller and younger subjects produced larger amounts of CO2 and consumed more O2 per unit of body weight. The use of a 5-min period when V̇CO2 varied by < 5% predicted a standard steady state.
Methods
Gas exchange measurements in mechanically ventilated subjects who underwent indirect calorimetry testing were analyzed from an existing repository of indirect calorimetry data. The study was approved by the institutional review board, and need for informed consent was waived. The methods for gas exchange measurement and data collection were described previously.10 In brief, the Vmax Encore (CareFusion, San Diego, California) was used to measure minute-to-minute V̇CO2 and V̇O2 for ∼30 min in each subject using non-dispersive infrared, galvanic cell, and mass flow sensors. Following calibration according to manufacturer specifications, connections were made on the inspiratory and expiratory limbs of the ventilator to measure inspired gas concentrations, expired gas concentrations, and exhaled gas flow. The calculation of V̇CO2 and V̇O2 was computed using dedicated software that employed the Haldane transformation.
In addition to indirect calorimetry data, subject demographics, endotracheal tube leak, diagnosis, and ICU stay at the time of indirect calorimetry were collected. Subjects were classified into general categories (medical or surgical) and then by specific disease process (respiratory, neurologic, neuromuscular, cardiac, neurosurgical, or other).
Description of Gas Exchange Measurement
Subjects who were not able to achieve steady state by the standard definition were excluded from the analysis. Data were analyzed using SPSS 21.0 (IBM, Armonk, New York) and Prism 5.04 (GraphPad Software, San Diego, California). Gas exchange values (V̇CO2 and V̇O2) were normalized to weight (mL/kg/min). Continuous values are expressed as median (interquartile range) if not normally distributed and as mean (95% confidence intervals) if normally distributed. Categorical variables are shown as number (percentage). Normalized gas exchange values (mL/kg/min) were plotted against the subjects' heights, and non-linear regression models were applied. The correlation coefficients were computed for each model to assess the goodness of fit. Analysis of variance was computed for the best performing regression, with a significance level set to P = .05. Additionally, V̇CO2 and V̇O2 values were summarized according to 3 distinct age groups (< 0.5 y, 0.5 to 8 y, and > 8 y). The differences in V̇CO2 and V̇O2 across age groups were assessed using one-way analysis of variance with Dunnett's multiple-comparison post-test (< 0.5 y group used as referent).
Steady-State V̇CO2 Comparison
Minute-to-minute V̇CO2 data were used to assess the agreement of various definitions for steady state. A total of 6 different definitions for steady state were tested based on the variation of V̇CO2 or of both V̇CO2 and minute ventilation over a 5-min interval for coefficient of variation (SD/mean) at 3 distinct thresholds (5%, 7.5%, and 10%). We compared the new definitions of steady state with the standard steady-state definition of a 5-min window during which V̇CO2 and V̇O2 vary by < 10%.9 A custom software macro was built to compute the steady-state definitions using Excel 14.0.6129.5000 (Microsoft, Redmond, Washington). Sensitivity, specificity, and the Fisher exact test were used to quantify the performance of the various V̇CO2-based steady-state definitions and to identify the one with the highest specificity.
Results
A total of 120 indirect calorimetry tests were completed in 92 mechanically ventilated subjects (with 28 repeat tests performed) during the eligibility period. A total of 33 (27.5%) tests, which were conducted in 22 subjects, failed to exhibit steady state. Steady-state measurements were achieved in 87 (72.5%) tests, which were conducted in 70 subjects, and were included in the analysis. The cohort included 35/70 males (50.0%) with median (interquartile range) age, height, weight, and FIO2 of 2.4 (0.68–13.9) y, 79.0 (57.5–118.2) cm, 12.6 (6.5–30.0) kg, and 0.4 (0.3–0.4), respectively. The median ICU stay at which point measurements were recorded was 46 (19–132) d. The median endotracheal tube leak was 4% (0–9%). A breakdown of subject diagnoses is provided in Table 1.
Description of Gas Exchange
The mean V̇CO2 and V̇O2 were 5.4 (95% CI 0.5) and 5.9 (95% CI 0.6) mL/kg/min for the entire cohort. A statistically significant relationship between normalized gas exchange and height was observed with power regression models: V̇CO2 = 115 × (height in cm)−0.71 (P < .001) (Fig. 1A) and V̇O2 = 130 × (height in cm)−0.72 (P < .001) (Fig. 1B). The correlation coefficients were both 0.61 for V̇CO2 and V̇O2. Table 2 shows gas exchange over distinct age groups. There were 14, 44, and 27 subjects in the age groups < 0.5 y, 0.5 to < 8 y, and ≥ 8 y, respectively. There were statistically significant differences in V̇CO2 among age groups: < 0.5 y compared with 0.5 to < 8 y (P < .05) and ≥ 8 y (P < .05). For V̇O2, differences were noted between the < 0.5 y group and ≥ 8 y group (P < .05).
Steady-State V̇CO2 Comparison
Minute-to-minute data from 47 subjects were utilized to compare the performance of the 6 steady-state definitions. The sensitivity, specificity, and Fisher exact test results are shown in Table 3. In this cohort, the modified steady-state definition, using a 5-min window for V̇CO2 values (with a coefficient of variation < 5%), predicted true steady-state conditions (obtained by the standard definition) with a sensitivity of 0.84 and a specificity of 1.0 (P < .01).
Discussion
Based on observations of our current cohort, gas exchange values significantly correlated with subject height and were also related to subject age. Both V̇CO2 and V̇O2 were produced at larger quantities per unit of body weight in younger subjects. Furthermore, the relationship between V̇CO2 and V̇O2 values (normalized to weight) and height could be described with a simple regression model. Furthermore, if height is not immediately available, one can obtain a reasonable estimate of expected values for V̇CO2 based on age groups. These findings provide a range of gas exchange values seen in a heterogeneous cohort and will aide in the interpretation of gas exchange values observed in the PICU. With the advent of stand-alone and ventilator-incorporated V̇CO2 monitors, a steady-state criterion to help assess reliable measurements is necessary. Standard steady-state conditions for indirect calorimetry testing in our cohort were predicted by a new definition (5-min period for V̇CO2 with a coefficient of variation < 5%) with high specificity and sensitivity.
Subject-to-subject differences may reflect variations in metabolic rate that could be attributed to level of activity, temperature, or stimulation at the time of the test. A comparative data set (as presented in this study) may enhance the ability to discern whether values for V̇CO2 and/or V̇O2 at the bedside are abnormally high or low. For example, if a 10-y-old patient is observed to have a V̇CO2 of 10.5 mL/kg/min (which corresponds to 3 times the expected value), careful assessment on the part of the clinician is likely indicated. Increases in V̇CO2 can be attributed to changes in clinical status (increased metabolism and increased CO2 production or changes in ventilation status). In contrast, decreases in V̇CO2 may indicate worsening pulmonary or cardiac performance, decreased consciousness, or inappropriate decreases in ventilator support. Institutional criteria may be developed to target those patients with abnormally high or low V̇CO2 for indirect calorimetry or other intensive assessment and monitoring.
In the present cohort, 27.5% of the tests did not achieve steady state. The recording of specific ventilator data and monitored parameters to determine those at risk for not achieving steady state was beyond the scope of our study, and therefore, we cannot say for certain why individual subjects failed to achieve steady state. However, it has been shown in adults that routine ICU care, ICU stay, disease state, and sedation can have a significant effect on absolute V̇O2 and energy expenditure levels, as well as minute-to-minute variation.12 In children, the levels of sedation and activity and the severity of disease are also associated with achievement of steady state.13,14 More specifically, increased variation in breathing frequency, ventilator circuit leak, and inspiratory time and higher peak inspiratory levels are associated with an inability to achieve steady state in critically ill, mechanically ventilated children and may result from changes in metabolism.10
The steady-state definition of 5 consecutive min of V̇CO2 data with a coefficient of variation < 5% has not previously been compared with the standard method, which uses a coefficient of variation < 10% for both V̇CO2 and V̇O2.9 The ability to accurately predict steady-state conditions using V̇CO2 alone may be useful when interpreting data from bedside CO2 monitors. Although the other definitions for steady-state V̇CO2 that were examined in our present study exhibited high sensitivities, the values for specificity were unacceptably low. In Table 3, it is clear that the definition incorporating minute ventilation in addition to V̇CO2 variability is quite acceptable, with a sensitivity and specificity of 0.82 and 1.00, respectively. The incorporation of minute ventilation variability into the steady-state definition does have the theoretical advantage of making the criteria more stringent and hence more accurate in a larger cohort. V̇O2 values contribute to the variations in metabolic state of critically ill patients; however, our study indicates that this contribution is relatively small during steady state and is likely closely related to variation in V̇CO2.
Values for gas exchange were strongly related to subject height and age. Lindahl et al15,16 demonstrated that a non-linear relationship between subject size and gas exchange exists in healthy anesthetized children, but included only children less than ∼25 kg in weight. The present cohort of mechanically ventilated children included a wider range of subject height and age and therefore may better reflect subjects admitted to the PICU. Our results confirm those of Lindahl et al,15,16 as they demonstrate that shorter and younger children have increased metabolism, as reflected by V̇CO2 and V̇O2 when indexed by size or weight, compared with older children.
Previous investigators reported values for V̇CO2 and V̇O2 for mechanically ventilated children in their PICU. Framson et al17 described a heterogeneous cohort of mechanically ventilated children in whom the mean V̇CO2 and V̇O2 were 4.3 and 5.2 mL/kg/min, respectively, with a mean age of 5.2 y and a mean height of 102 cm. In a survey of 100 mechanically ventilated children, White et al18 reported a mean V̇CO2 and V̇O2 of 5.3 and 6.3 mL/kg/min, respectively, for a cohort with a mean age of 4.5 y and a mean height of ∼97 cm. Furthermore, in a slightly younger population with a mean age of 4.2 y, Martinez et al19 reported a mean V̇CO2 and V̇O2 of 4.8 and 6.1 mL/kg/min, respectively, for a cohort of mechanically ventilated children recently admitted to the PICU. These findings suggest that the values from our study are largely in agreement with values reported elsewhere. However, none of the aforementioned studies reported data normalized to subject size. Our models should be tested in other data sets.
There are important limitations to this study that must be carefully considered. First, the results reflect findings from a single institution and represent a heterogeneous collection of subject sizes and disease states. However, the population does reflect a patient population that is likely to be encountered in the PICU. Second, we used a single device to make gas exchange measurements. Reports suggest that gas exchange data do not necessarily agree among approved devices, with various factors contributing to the disagreement.20 We chose to utilize a single device to ensure that device-device agreement was not a confounding factor in our study. However, device agreement and individual accuracy are important considerations, and extrapolation of the present findings to devices other than that tested must be done with caution. Furthermore, the device used in this investigation was shown in an in vitro simulation of child metabolism to be accurate and reliable.21 Third, the time points of data collection were not standardized among the cohort, and accordingly, subjects may have been in different phases of their disease. Because the median duration of ICU stay at which indirect calorimetry measurements were taken was 46 d, our results reflect subjects in a non-acute phase of their disease. However, because the hypermetabolic response noted in critically ill adult patients is not apparent in pediatric critically ill children, differences in stay at the time of indirect calorimetry should not have a significant effect on gas exchange values.17 Indeed, it has been shown that critically ill, mechanically ventilated children do not tend to exhibit predictable changes in gas exchange as a function of time or disease phase in the ICU.17 It is possible that certain diseases may exhibit different findings for gas exchange values compared with other diseases, but the number of patients required to demonstrate these differences was prohibitively large. Because it was the aim of this study to describe gas exchange values for the cohort that could yield a single tool for clinicians, we decided not to pursue further analyses based on subcategorized diseases.
Future studies should investigate the feasibility of incorporating steady-state V̇CO2 methods into available ventilators and CO2 monitors to ensure the quality and reliability of minute-to-minute V̇CO2 data available at the bedside. Software applications that include steady-state criteria thresholds, detection of outliers, and trend views will allow clinical decision making based on these data. Devices should also include onboard calculation of gas exchange normalized to body weight (mL/kg/min). With the increasing availability of bedside gas exchange monitors and monitors incorporated into ventilators, our observations will help with interpretation of the continuous data and allow them to be incorporated into clinical decision making.
Conclusions
In a cohort of mechanically ventilated children, we observed a large range of V̇CO2 and V̇O2 measurements. Gas exchange values were strongly related to height and age; smaller and younger patients tended to produce larger amounts of CO2 per unit of body weight. A new steady-state definition of a 5-min period for V̇CO2 with a coefficient of variation < 5% was in reasonable agreement with the standard definition and may be used to determine the validity of continuous V̇CO2 measurements at the bedside. Our observations may facilitate greater utility of V̇CO2 at the bedside in the PICU and thereby extend the benefits of metabolic monitoring to a larger group of patients when indirect calorimetry is not available.